Sets out a flame atomic absorption method for the determination of manganese in natural and waste waters in the range of 0.005 mg/L to 5 mg/L. Manganese in the range of 0.05 mg/L to 5 mg/L can be determined by direct aspiration while the range 0.005 mg/L to 0.05 mg/L requires a pre-concentration step.